Test piece

Name of test piece/s ATENA SYSTEM FOR HOSPITAL FALSE CEILING
Description Name of ceiling panel: Atena Syncro Evo; name of lighting system: Taurus Evo (unplugged)
Project number AT 2504-1620
Created on 09/25/2025
Valid until 08/15/2030
Test type Particle emission (ISO 14644-1, -14)
Classfication Suitability for air cleanliness class: (ISO-Class):3
Particle emission (ISO 14644-1, -14)

Particle emission (ISO 14644-1, -14)

Test parameter(s)

Air Cleanliness Class

Structure-borne noise = approx. 50 Hz

3

Overall result

3


It must be pointed out, that according to ISO 14644-1 cleanrooms classes 1 to 5 have a high filter occupancy, with the result that large-surface ceiling/lighting systems cannot be used in some cases. Cleanrooms with a horizontal displacement flow form an exception to this.
The test result may be influenced by the surrounding ceiling system, in particular the material pairing between the light and ceiling frame, as well as other assembly accessories. Particle emission behavior should be re-assessed in the respective assembly situation. The cut edges/back are made of very porous material.

Please note: Transport damages, incorrect installation, aging behavior, corrosion etc. can influence the test result.